IviDigitizer Trigger Holdoff
IviDigitizerTriggerHoldoff Property
Data Type | Access | Applies To | Coercion | High Level VIs |
---|---|---|---|---|
R/W | N/A | Down | Configure Trigger Holdoff |
Description
Specifies the length of time the digitizer waits after it detects a trigger until the digitizer enables the trigger subsystem to detect another trigger. The units are seconds. The Trigger Holdoff property affects instrument operation only when the digitizer requires multiple acquisitions to build a complete waveform. If Trigger Holdoff and PretriggerSamples are are both non-zero, then both conditions must be satisfied before the digitizer will accept a trigger. Note: Many digitizers have a small, non-zero value as the minimum value for this property. To configure the instrument to use the shortest trigger hold-off, the user can specify a value of zero for this property. Therefore, the IVI Class-Compliant specific driver shall coerce any value between zero and the minimum value to the minimum value. No other coercion is allowed on this property.