Admin Production ni-theme
Current Publication

IviDigitizer Trigger Source List

IVI Compliance Package

IviDigitizer Trigger Source List

IviDigitizerMultiTrigger Property

Data Type Access Applies To Coercion High Level VIs
R/W N/A None Configure Multi Trigger

Description

A comma separated list of source names to be used in a multi trigger scenario. This property only affects instrument behavior when the Trigger Source Operator property is set to AND or OR. When Trigger Source Operator is set to AND, the trigger conditions associated with each source in this list must be simultaneously satisfied in order to trigger the digitizer. When Trigger Source Operator is set to OR, the first trigger source in the list that satisfies its trigger conditions will trigger the digitizer. Any valid name used for the Trigger Source property may be used in this list. A trigger source may appear only once in the list. If a name in the list is not recognized, the driver returns the Unknown Channel Name error.